FUJIFILM Life Science
Products
BulletLAS-4000
BulletLAS-4000mini
BulletFLA-5100
BulletFLA-7000
BulletFLA-9000
Features
Imaging Samples
System Spec
BulletBAS-5000
BulletAP-3000
BulletImaging Plates
BulletAccessories
BulletMore Products
BulletReagents
FLA-8000

Image Reading Software

Windows®

Dimensions and Weights

850(W) x 730(D) x 480(H)mm, 110kg

Lasers

Two fixed lasers for fluorescent dye excitation and reading radioisotopic image            

532nm SHG
635nm LD

One optional laser for fluorescent dye excitation            
473nm SHG

Nuclides for RI Image Detection
14C, 32P, 33P, 35S, 125I (the numbers in front should be small in front, i.e. 14th)

Dynamic Range

Five orders of magnitude

Bit Depth

16 bits (gray scale: 65, 536)

Filters

Three fixed filters for modularity:            
530nm
570nm
675nm

One optional filter chosen by user:            
25mmø t2-t12mm

NOTE: In case of IP detection, this position is occupied by a B390 filter for photo-stimulated luminescence.

Modularity

The most suitable FLA-8000 can be upgraded depending on your purpose by selecting any optional component.

Recommended Uses/Applications

RI Image:            
Radioisotope-labeled macroarray membrane
In situ hybridization
Radioisotope-labeled electrophoresis gel
Radioisotope-labeled rat brain autoradiography

Fluorescent Image:            
Fluorescent dye stained gel (mini-gel)
Fluorescent dye labeled electrophoresis
FISH on glass slide
GFP image for protein analysis
*No license is granted for the use in combination with fluorescent microarray glass slide.

Sample Size and Scanning Size

 

Sample Size (mm)

Scanning Size (mm)

Pixel Size (micron)

Slide Carrier

5 glass slides

5x(74x23)

5/10/20

IP Carrier

85x127 IP

85x127

10/20/50/100

Gel Carrier

2 gel sheets (51x59)

2x(50x58)

100

Representative Fluorescent Dyes

473nm SHG Laser:            
SYPRO® Orange,
SYPRO® Ruby,
SYBR® Green I,
FITC

532nm SHG Laser:            
Cy3™,
Rhodamine,
SYPRO® Red

635nm Laser:            
Cy5™

Optional Components

473nm SHG laser
Sample carriers
(IP Carrier, Slide Carrier, Gel Carrier)
IP cassette
Filter (B390 for IP reading)
BAS-SR0813 IP
IP Eraser 3

Bottom Nav